Reliable in all conditions
Where there is a need to monitor not only groundwater levels but also saltwater intrusion, injected
wastewater, or contamination from chemical discharges and landfill sites, the ceramic-casing on the
CTD-Diver is the instrument of choice. In addition to the pressure and temperature sensor, the CTD-Diver
has a four-electrode conductivity sensor for reading conductivity
across an expanded measurement range
(0-120 mS/cm). There are two options for measuring conductivity: displaying measured conductivity or
specific conductivity at 25 °C.
The CTD-Diver accurately records up to 48.000 measurements of groundwater levels, temperature and
conductivity with corresponding date and time.
The one-eye communication makes this compatible
with the Diver-Suite.
Highlights
Hermetically sealed in ceramic housing
- Various measurement methods: fixed, event dependent, averaging and pumping test
- Robust corrosion resistant housing
- 10 years battery life
Conductivity specifications
User adjustable range 10 ?S/cm to 120 mS/cm
Typical accuracy ± 1 % of reading
Resolution ± 0.1 % of reading
Pressure specifications
Type 11.11.59.01 11.11.59.02 11.11.59.03
Range 10 m H2O 50 m H2O 100 m H2O
- Typ. accuracy :
±0.5 cm H2O ±2.5 cm H2O ±5.0 cm H2
|
General specifications
Dimensions Ø 18-22 mm × 135 mm
Memory 48.000 measurements
Wetted parts
housing
• ceramic (ZrO2)
conductivity sensor housing
• ceramic (ZrO2)
conductivity sensor
• platinum electrodes on
ceramic (Al2O3) carrier
o-rings
• fluorocarbon rubber (FKM)
pressure sensor
• ceramic (Al2O3)
cap / nose cone
• Akulon (PAG 30%)
Battery life 10 years
Sample interval 1 second to 99 hours
Mass 95 grams
Temperature specifications
Range / compensated -20 °C to 80 °C / 0 °C to 50 °C
Typical accuracy ±0.1 °C
Resolution 0.01 °C

Download the latest